TYPICAL PRODUCT PARAMETERS

TYPICAL PRODUCT PARAMETERS

Electrical Parameters

 

MA90Z

 

MA95Z

 

 

 

 

INPUT VOLTAGE AC:

 

90 V

 

 

 

INPUT FREQUENCY:

 

60 Hz

 

 

 

 

CURRENT:

1 A

 

2 A

 

 

 

 

MAXIMUM POWER CONSUMPTION:

30 ± 10 W

 

150 ± 10 W

 

 

 

 

CIRCUIT PROTECTION:

Dual fused,

 

Dual fused,

 

250 V,

 

250 V,

 

1 A fast blow fuses

 

5 A fast blow fuses

 

 

 

 

MODE OF OPERATION:

 

Continuous

 

 

 

 

Performance Parameters

 

MA90Z

 

MA95Z

 

 

 

 

WEIGHT CAPACITY

 

 

 

STANDARD MATTRESS:

350 lbs

 

BARIATRIC MATTRESS:

1000 lbs

 

 

 

 

PRESSURE ZONE:

2

 

 

 

 

MAXIMUM FLOW:

50 ± 15 LPM

 

1275 ± 100 LPM

 

 

 

MAXIMUM PRESSURE:

35 ± 5 mmHg

 

 

 

MAXIMUM FLOW TIMER:

30 minutes

 

15 minutes

 

 

 

 

SUPPORT SURFACE INFLATION TIME:

5 - 10 minutes

 

25 - 60 seconds

 

 

 

 

PATIENT COMFORT CONTROL PRESSURES:

 

 

 

SOFT PRESSURE:

8 ± 4 mmHg

FIRM PRESSURE:

32 ± 4 mmHg

 

 

 

ROTATION TIME:

10, 20, 30, 60 minutes

 

5 minutes - 4 hrs

 

 

 

 

ROTATION ANGLE:

0 - 30° ± 5°

 

0 - 40° ± 5°

 

 

 

PATIENT CONTACT:

Control unit and mattress have

 

Latex-Free components

 

 

 

 

Part No 1148139