CV-M300

7.3. Jumper Settings

Cautions on Jumper Setting.

Before making any mode or jumper setting turn the power OFF.

Jumpers for mode setting are found on the boards PK8041 & PK8042.

The following modes are available with jumper setting:

Input/Output Mode of HD/VD signal. (HD/VD input is factory setting)

Edge Pre-select Mode. (Factory setting)

Pulse Width Control Mode

Start/Stop Trigger Mode

Long Time Exposure Mode

Set the jumpers according to the list below in 7.3.1. and 7.3.2.

 

 

7.3. . Jumper onHD/VDPK8042in/outBoard

 

Externaltriggermodes

 

 

 

The jumper positions are shown in “7.4.1. Board 8041 Side A” and “7.4.2. Board 8042 Side A”.

Jumper

HD/VDExt.

HD/VDInt.

Edgeselectpre-

widthPulse

Start/stp

Longexposetime

 

Remarks

Switch SW2 on PK8041 is shown in “7.2. SW2 on PK8041 Board”.

FactorysetSHORT,PWCopen

JP14

Short

Short

 

Short

Open

Short

Short

 

input*)

output

 

 

control

trigger

 

 

 

7.3.2. Jumper onHD/VDPK8041in/outBoard

 

Externaltriggermodes

Longtime

 

Remarks

Jumper

HD/VDExt.

HD/VDInt.

Edgepre-

widthPulse

Start/stop

 

*) Factory Setting

 

 

selct

 

 

expose

 

 

6

input*)

output

 

control

trigger

Longtimeintegration

87

Short

Short

 

Short

Short

Short

Short

HDinp

t/outputonpin7

JP901

Short

 

Open

Short

VDWEN

SW2-21

ON

OFF

 

ON

ON

ON

ON

HDVinput/output75Ohmterm.FactorysetON

JP123

Open

Open-

 

Open-

Open-

Short-

Open-

Pixelclock .FactorysetOPEN

*) Factory Setting

7.3.3. Pixel Clock Output

To use pixel clock output (TTL level: 4.0 V), make the jumper JP13 on PK8041 board shorted-circuited. The jumper position is shown in “7.4.1. Board PK8041 Side A”.

Pixel clock pulse will be output from pin no.9 of 12-pin multi-connector.

Cautions for Pixel Clock Output.

When the pixel clock is enabled, it may cause interference with external equipment if it not properly shielded.

- 16 -

Page 17
Image 17
JAI CV-M300 operation manual Jumper Settings, Pixel Clock Output