5.Attach the appropriate detector adapter to the detector on the front of the meter.
6.Connect the output connector of the measurement jumper to the detector adapter. This end of the measurement jumper is
7.Press the Power key to set the meter to Power mode. When in Power mode, the meter displays P at the beginning of the reading, and no “rel” is displayed, for example, P =
8.Press and hold the Power key until the display on the meter clears.
9.Release the Power key. The display reads 0.00 dB, and the loss of the measurement jumper is subtracted from the DUT loss or power measurements to be made (Figure 7).
Figure 7: Loss and Power Measurements
For loss and power measurements below
10.Ensure that the detector is covered, for example, that the detector cap is on or the output connector is connected to the detector. This dark measurement current is subtracted from future loss and power measurements.
11.Press the Store ID key. The key lamp lights to indicate that the
Measuring Power
In Power mode, absolute power measurements are displayed in dBm. To make a power measurement of a DUT after performing the setup steps in the Setting Up the Meter for Loss and Power Measurements section:
1.Press the Power key to set the meter to Power mode.
2.Press the λ key to select the wavelength at which the measurement is to be done.
3.Clean the output connector of the measurement jumper and the input connector of the DUT, and connect the two connectors.
10112341 Rev 002 | Page 28 of 39 |