EMC Guidance
Guidance and manufacture’s declaration - electromagnetic emissions
The Oxygen Saturation Monitor Model
Emissions test | Compliance | Electriomagnetic environment - guidance | |
RF emissions |
| The Model | |
CISPR 11 | Group 1 | its internal function. Therefore its RF emissions are | |
|
| very low and are not likely to cause any interference | |
|
| in nearby electromagnetic equipment. | |
RF emissions | Class B | The Model | |
CISPR 11 | establishments, including domestic establishments | ||
| |||
|
| and those directly connected to the public | |
Harmonic emissions |
| ||
Not applicable | |||
| buildings used for domestic purposes. | ||
Voltage fluctuations/ | Not applicable |
| |
flicker emissions |
| ||
|
|
Appendix
Guidance and manufacture’s declaration - electromagnetic immunity
The Oxygen Saturation Monitor Model
Immunity test | IEC 60601 | Compliance | Electromagnetic environment - |
| Test level | level | guidance |
|
|
|
|
Electromagnetic | ±6 kV contact | ±6 kV contact | Floors should be wood, concrete or |
Discharge (ESD) | ±8 kV air | ±8 kV air | ceramic tile, if fllors are covered with |
|
| synthetic materisl, the relative humidity | |
|
| should be at least 30%. | |
|
|
| |
Electrical fast | Not applicable |
|
|
Transient/burst |
|
|
|
|
|
| |
Surge | Not applicable |
|
|
|
|
| |
|
|
|
|
Voltage dips, short | Not applicable |
|
|
Interruptions and |
|
|
|
Voltage variations |
|
|
|
on power supply |
|
|
|
Input lines |
|
|
|
|
|
| |
|
|
|
|
Power frequency | 3 A/m | 3 A/m | Power frequency magnetic fields |
(50/60Hz) magnetic |
|
| should be at level characteristic of a |
field |
|
| typical location in a typical commercial |
|
| or hospital environment. | |
|
|
| |
|
|
|
|
32