measurement taken). You can select a particular measurement by using a longer press on the PLUS - key {A, 3} or a longer press on the MINUS - key {A, 13}. When selected the measurement blinks. Once selected and blinking, the measurement can be transmitted using the BLUETOOTH® - key {A, 14}. The same selection process can be used for areas, volumes, indirect measurements and room dimen- sions.
Transmission of user-entered measurements
A measurement on the display can be changed if you wish. By pressing the EQUAL - key {A, 5} the value blinks and can then be changed using the PLUS - key {A, 3} and the MINUS - key {A, 13}. Pressing the keys for longer increases the rate of change. Pressing the EQUAL - key {A, 5} again confirms the change and the value can be transmitted as normal.
)TheNOTEfree software supplied is designed to provide the user with basic functionality. Leica Geosystems provides no warranty for this free soft- ware and offers no support for it. Leica Geosystems accepts no liability whatsoever arising from the use of the free software and is not obliged to provide corrections nor to develop updates or upgrades.
You can find numerous commercial vendors of software for a wide range of applications on our home- page.
Appendix
GB
Message Codes
All message codes will be displayed with either "InFo" or "Error".
The following mistakes can be corrected:
InFo | Cause | Remedy |
204 | Calculation error | Repeat procedure |
206 | No endpiece detec- | Attach the endpiece |
| tion | properly. If error still |
|
| occurs, replace the |
|
| endpiece. |
240 | Error in data trans- | Repeat process |
| mission |
|
245 | 2nd key pressed | Establish BLUE- |
| although no BLUE- | TOOTH® connection |
| TOOTH® connection |
|
| exists. |
|
252 | Temperature too | Cool down instrument |
| high |
|
253 | Temperatur too low | Warm up instrument |
255 | Receiver signal too | Use target plate |
| weak, measurement |
|
| time too long, |
|
| distance > 100 m |
|
256 | Received signal too | Use target plate |
| powerful | (grey side) |
257 | Wrong measure- | Use target plate |
| ment, ambient | (brown side) |
| brightness too high |
|
260 | Laser beam inter- | Repeat measurement |
| rupted |
|
Leica DISTO™ A6 1.0.0 gb | 9 | Appendix |
|
|
|