18GETTING YOUR InDuo™ SYSTEM READY FOR USE

Apply control solution

Shake the control solution vial well.

Remove the cap.

b• Squeeze the vial and discard the first drop and

wipe off the dispenser tip with a clean tissue to

obtain an accurate result.

• Squeeze the vial again to get a small hanging drop.

• Apply the hanging drop b to the top edge area of the OneTouch® Ultra Test Strip (where it meets the narrow channel) until the drop is drawn into the OneTouch® Ultra Test Strip, the confirmation window is filled, and the meter display begins to count down.

Read test results

c • The meter will count down from N to A and the control solution test result will appear c.

• Compare the result with the range printed on the OneTouch® Ultra Test Strip vial label. The result should fall within this range.