50INJECTING YOUR INSULIN
20
InDuo™ insulin doser function check
If you think that the InDuo™ insulin doser is not working properly, follow this procedure:
•Screw on a new NovoFine® needle, as described on page 41.
•Prime the insulin doser as described on page 42.
•Put the outer needle cap over the needle.
•Dispense 20 units into the outer needle cap.
The insulin should fill the lower part of the outer needle cap.
•If too much or too little insulin is released, repeat the test.
•If the InDuo™ insulin doser again releases too much or too little insulin, do not use your insulin doser. Do not try to repair a faulty insulin doser. Call the InDuo™ Customer Care Line at
1 888