By using the variant 504-02 relay 2 will be controlled by the MJB variant 505/506 with switch input.

Alarm 1 = Switch input S2 (OoBD alarm) / Alarm 2 = Switch input S1

When using this combination, is it important to have the MJB 505/506 connected to the system all the time. If it is not connected, the Gateway MJB will see it as an alarm and the relay will be activated. The Gateway MJB is “scanning” the OpenBus™ system every minute for an alarm.

Features:

Add-on value for hospital & care bed

Compact and functional design which allows easy mounting

2 connections, Port1 OpenBus™ connection / Port 5 Gateway

Colour: Light grey RAL 7035

Protection class IPX6

Connection to OpenBus™ control box via modular plug cable, 6 wires

Modular plug to Open end, see precautions.

Options (special article):

Customised SW:

It is possible to define the OpenBus™ signals to be used for activating the relays and / or change the alarm indicator, so that it fits to the customers system. Please contact your Sales Engineer at LINAK A/S for further information.

Usage:

Compatible with all LINAK OpenBus™ Control Boxes, light & full Contact rating :

• Max continuous current :

= 1A

• Maximum switching voltage :

= 125V (DC) / 125V (AC)

• Maximum switching capacity :

= 24W (DC) / 62,5VA (AC)

• Contact resistance :

< 100 mΩ

Current consumption:

 

• V bus 8V :

= 9 mA

• V permanent 40V :

= 14 mA

Approvals: LINAK and UL International DEMKO A/S are currently in the process of certifying the MJBX0X, re. UL60601 and EN60601-1