Linak MJB user manual Relay, Example of closed loop, Example of Open Loop

Models: MJB

1 24
Download 24 pages 31.94 Kb
Page 12
Image 12

Relay 1

CB power mode

Alarm level

Relay state

 

 

 

Mains unplugged, CB is in power down or the OpenBus™ is not

Alarm

NC\*

running (Clock/ data is missing)

 

 

 

 

 

On mains or battery at “wake up”

Bed Alarm\**

NC

 

 

 

 

No Alarm

NO

 

 

 

 

 

 

Relay 2

 

 

 

 

 

CB power mode

Alarm level

Relay state

 

 

 

Mains unplugged or CB is in power down

No Alarm

NC

 

 

 

On mains or battery at “wake up”

Bed Alarm\**

NO

 

 

 

 

No Alarm

NC

 

 

 

\* When the mains is unplugged or CB is in power down will the relay shift to NC state and generate alarm.

\** Alarm generated by nurse call or bed alarms

Example of closed loop:

3rd. party site

In this state no Alarm

Example of Open Loop:

3rd. party site

In this state no Alarm

12

Page 12
Image 12
Linak MJB user manual Relay, Example of closed loop, Example of Open Loop