565DG/565TG Ortho-Biotic Preferred Care®Recliners Operating Instructions
8
Figure 13

RECLINER POSITIONING MECHANISM

TROUBLE-SHOOTING

ADJUSTMENT INSTRUCTIONS FOR POSITION
MECHANISM
NOTE:Never lubricate the position
mechanism assembly.
1. Turn the position adjustment handle down to
the release position, and remove clevis pin
and wire ring from top of assembly.
(See fig. 13).
2. Slide positioning plate up, and turn square
washer clockwise until snug (See fig. 13).
3. Slide the positioning plate down over square
washer. Replace clevis pin and wire ring.
4. The position mechanism's handle should
engage at about the halt way point. If more
tension is desired, repeat the above procedure.
Clevis ring and
wire ring
Positioning plate
Square washer
Slide