Manley Labs 120 Watt Monoblock Amplifier owner manual Bias Procedures

Models: 120 Watt Monoblock Amplifier

1 10
Download 10 pages 28.49 Kb
Page 6
Image 6
BIAS PROCEDURES

BIAS PROCEDURES

The Manley 120 Watt Amplfier uses a variable "fixed bias" system that requires very little attention. If you wish to adjust or check the bias, complete the following steps. For this procedure you will need a multimeter to read millivolts (less than 2 volt range) and a small flat screwdriver.

1.Looking at the front section of the top plate of the amplifier, you can see 2 red tip jacks and 1 black tip jack. To measure the current draw of each output tube place the positive probe (red) into the red tip jack for the tube you wish to measure and the negative probe (black) in the black (GROUND) tip jack and set the meter to read 'millivolts'.

2.Each of the two (2) output tubes can have its own bias adjusted by altering its own bias adjust pot. The numbering sequence for the 2 output tubes runs from left to right, as do the corresponding tip jacks (measurement points). The bias adjust pots are located on the top side of the amplifier, therefore it will be NOT be necessary to remove the bottom cover to changing the bias. They also read left to right.

3.The first step to setting the bias is to turn on the amplifier and insure that there is zero signal input. Mute your preamplifier. Don't play tunes. Also, you MUST leave your loadspeakers connected as these provide an ideal load on the output. Never operate your amplifiers without a suitable load or speakers attatched. It is recommended that you leave the amplifier on long enough to ensure that the tubes have reached their stable current draw.

5.Set your meter to read VOLTS DC. You will be measuring less than 1 volt so set your meter for a low range if it is not an autoranging meter. Place the probes in the first red tip jack (negative probes remains in the black tip jack for whole biasing procedure) and adjust the first bias trim pot located on the left side top deck of the amplifier until you measure approx. 300 mVDC (0.300 VDC). Since you are measuring across a 10 ohm cathode resistor, this would correspond to a 30 mA bias current. Settings between 275 mV and 300 mV are perfectly OK but try to have each tube the same.

6.If you are unable to bring the bias voltage down to 300 mV on any tube then we recommend that the tube be carefully watched or replaced.

7.Follow step five with the other tube. Then go back and recheck again. Check and set the bias on your amplifiers every day for the first week or any time you retube. Then check every week, then every month. A red glowing area on the side of the tube usually indicates it is drawing way too much current. If you see a red spot on the side of the tube, check and reset your bias immediately.

Page 6
Image 6
Manley Labs 120 Watt Monoblock Amplifier owner manual Bias Procedures