Melissa DVP3980 manual Error Messages, Error message Meaning Cause, Solution

Models: DVP3980

1 43
Download 43 pages 41.83 Kb
Page 30
Image 30

630012_IM6_rev 23/10/06 10:42 Side 30

ERROR MESSAGES

The apparatus registers any problems and shows the following error messages on the display:

Error message

Meaning

Cause

ER-1

Error during

1.

The armband is not correctly attached.

 

pumping

2.

The armband is broken.

 

 

3.

The pump or valves are defective.

ER-2

Disturbances or

1.

The user moves or speaks

 

noise during readings

during the reading.

 

 

2.

The user is too nervous.

reading taken by a doctor.

 

 

ER-3

Error while emptying

1.

The valves are defective.

 

the armband

2.

The armband is not correctly attached.

ER-4

Abnormal reading

The user may have a heart problem.

HI

The pressure in the

The air tube is blocked, the pump

 

armband is too high.

mechanism is defective or the pressure

 

 

monitor is defective.

Battery symbol

Flat batteries

The batteries are almost dead.

Solution

1.Attach the armband correctly.

2.Contact the retailer.

3.Contact the retailer.

1.The user must sit still and avoid talking during the reading.

2.Have a blood pressure

1.Contact the retailer.

2.Attach the armband correctly.

Consult a doctor.

Try to take another reading, or contact the retailer.

Replace the batteries.

30

Page 30
Image 30
Melissa DVP3980 manual Error Messages, Error message Meaning Cause, Solution