 51
SCALE OPERATION  
CALIBRATION  
Bar Code Calibration Procedure without Remote Display 
11.  Remove half of the load and center the remaining load.  The white LEDs will blink four times periodically.  Wait for scale 
stability, and then press the beeper volume switch once.  The beeper will beep 4 times indicating that the Cal 4 value has 
been stored. 
Figure 58. 
12.  Remove the remaining half load from the scale.  The white LEDs will blink five times periodically.  Wait for scale stability, 
and then press the beeper volume switch once. The beeper will beep 5 times indicating that the Cal 5 value has been 
stored. 
Figure 59. 
13.  Calibration is now complete.  The scanner will automatically restart, and beep one time, in 5 seconds. 
14.  In order to use the scale in the normal operating mode, the unit must be powered down and the scale program switch 
returned to the run position.  
Figure 60. 
See Power Source caution statement located on page 11 of this manual. 
15.  The unit’s calibration must now be verified as required by state and/or local Weights and Measures regulations (starting 
on page 52).