Chapter 4 Analog Input
NI 6238/6239 User Manual 4-4 ni.com

Figure 4-2. Analog Current Input Connection Method 1

Method 2

Method 2, shown in Figure4-3, ties th e AI – input to AI GND. When

measuring current up to 20 mA, this type of connection ensures that the

voltage level on both the positive and negative side are within the

common-mode input range for NI 6238/6239 devices.

Figure 4-3. Analog Current Input Connection Method 2

+
AI +
AI –
Isolation
Barrier
AI GND
V
AI GND
+
+
AI +
AI –
Isolation
Barrier
AI GND