Chapter 4 Analog Input
© National Instruments Corporation 4-21 NI 6238/6239 User Manual
Figure 4-13. Single External Signal Driving ai/SampleClock and ai/ConvertClock
Simultaneously
AI Convert Clock Timebase SignalThe AI Convert Clock Timebase (ai/ConvertClockTimebase) signal is
divided down to provide on of the possible sources for ai/ConvertClock.
Use one of the following signals as the source of
ai/ConvertClockTimebase:
• ai/SampleClockTimebase
• 20 MHz Timebase
ai/ConvertClockTimebase is not available as an output on the I/O
connector.
AI Hold Complete Event SignalThe AI Hold Complete Event (ai/HoldCompleteEvent) signal generates a
pulse after each A/D conversion begins. You can route
ai/HoldCompleteEvent out to any output PFI <6..9> or RTSI <0..7>
terminal.
The polarity of ai/HoldCompleteEvent is software-selectable, but is
typically configured so that a low-to-high leading edge can clock external
AI multiplexers indicating when the input signal has been sampled and can
be removed.
• One External Signal Driving Both Clocks
ai/SampleClock
ai/ConvertClock
Sample #1 Sample #2 Sample #3
123012301…0
Channel Measured