Chapter 1 Introduction
© National Instruments Corporation 1-9 NI 781xR User Manual
Operate the NI781 xR at or below the measurement category1 listed in the 
Environmental section of AppendixA, Specifications. Measurement 
circuits are subjected to working voltages2 and transient stresses 
(overvoltage) from the circuit towhich they are connected during 
measurement or test. Measurement categories establish standard impulse 
withstand voltage levels that commonly occur in electrical distribution 
systems. The following list describes installation categories:
•Measurement Category I—Measurements performed on circuits not 
directly connected to the electrical distribution system referred to as 
MAINS3 voltage. This category is for measurements of voltages from 
specially protected secondary circuits. Such voltage measurements 
include signal levels, special equipment, limited-energy parts of 
equipment, circuits powered by regulated low-voltage sources, and 
electronics.
•Measurement Category II—Measurements performed on circuits 
directly connected to the electrical distribution system. This category 
refers to local-level electrical distribution, such as that provided by a 
standard wall outlet (for example, 115 V for U.S. or 230 V for Europe). 
Examples of Measurement Category II are measurements performed 
on household appliances, portable tools, and similar products.
•Measurement Category III—Measurements performed in the 
building installation at the distribution level. This category refers to 
measurements on hard-wired equipment such as equipment in fixed 
installations, distribution boards, and circuit breakers. Other examples 
are wiring, including cables, bus-bars, junction boxes, switches, 
socket-outlets in the fixed installation, and stationary motors with 
permanent connections to fixed installations.
•Measurement Category IV—Measurements performed at the 
primary electrical supply installation (<1,000 V). Examples include 
electricity meters and measurements on primary overcurrent 
protection devices and on ripple control units.
1   Measurement categories, also referred to as installation categories, are defined in electrical safety standard IEC 61010-1.
2   Working voltage is the highest rms value of an AC or DC voltage that can occur across any particular insulation.
3   MAINS is defined as a hazardous live electrical supply system that powers equipment. Suitably rated measuring circuits may 
be connected to the MAINS for measuring purposes.