© National Instruments Corp. 17 NI 9235/9236

Stopband

The filter significantly attenuates all signals above the stopband
frequency. The primary goal of the filter is to prevent aliasing.
Therefore, the stopband frequency scales precisely with the data
rate. The stopband rejection is the minimum amount of attenuation
applied by the filter to all signals with frequencies within the
stopband.

Alias-Free Bandwidth

Any signal that appears in the alias-free bandwidth of the
NI9235/9236 is not an aliased artifact of signals at a higher
frequency. The alias-free bandwidth is defined by the ability of the
filter to reject frequencies above the stopband frequency and equals
the data rate minus the stopband frequency.
Understanding NI 9235/9236 Data Rates
The frequency of a master timebase (fM) controls the data rate (fs)
of the NI 9235/9236. The NI 9235/9236 includes an internal master
timebase with a frequency of 12.8 MHz, but the module also can
accept an external master timebase or export its own master
timebase. To synchronize the data rate of an NI 9235/9236 with
other modules that use master timebases to control sampling, all of