NI 9235/9236 26 ni.com
Crosstalk (fin = 1 kHz)...................... –100 dB
Common-mode voltage,
all signals to earth ground.................±60 VDC
CMRR (fin = 0 to 60 Hz)
NI 9235.......................................120 dB
NI 9236.......................................110 dB
MTBF............................................... 566,796 hours at 25 °C;
Bellcore Issue 2, Method 1,
Case 3, Limited Part Stress
Method
Note Contact NI for Bellcore MTBF specifications
atother temperatures or for MIL-HDB K-217F
specifications.