NI 9235/9236  30 ni.com
Isolation
Channel-to-channel.................... None
Channel-to-earth ground
Continuous........................... 60 VDC, 
Measurement Category I
Withstand............................. 1,000 Vrms, verified by a 5 s 
dielectric withstand test 
Measurement Category I is for measurements performed on 
circuits not directly connected to the electrical distribution system 
referred to as MAINS voltage. MAINS is a hazardous live electrical 
supply system that powers equipment. This category is for 
measurements of voltages from specially protected secondary 
circuits. Such voltage measurements include signal levels, special 
equipment, limited-energy parts of equipment, circuits powered by 
regulated low-voltage sources, and electronics.
Caution Do not connect the NI 9235/9236 to signals or 
use for measurements within Measurement CategoriesII, 