© National Instruments Corp. 21 NI 9235/9236
Data rate range (fs) using internal master timebase
Minimum....................................794 S/s
Maximum.................................. .10 kS/s
Data rate range (fs) using external master timebase
Minimum....................................195.3125 S/s
Maximum................................... 10.547 kS/s
Data rates1 (fs)..................................... , n = {2; 4, 5, ..., 63}
Full-scale range.................................±29.4 mV /V
(+62,500 με/–55,500 με)
Scaling coefficient............................ 3.5062 nV/V per LSB
Overvoltage protection
between any two terminals............ ...±30 V
1 The data rate must remain within the appropriate data rate range. Refer to the
Understanding NI 9235/9236 Data Rates section for more information.
f
M256÷
n
---------------------