© National Instruments Corporation 25 NI PXI/PCI-5124 Specifications
EnvironmentNI PXI-5124
Note To ensure that the NI PXI-5124 cools effectively, follow the guidelines in the
Maintain Forced Air Cooling Note to Users included in the NI PXI-5124 kit. The
NI PXI-5124 is intended for indoor use only.
Specification Value Comments
Operating
Temperature
0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-1000/B or
PXI-101x chassis
Meets IEC-60068-2-1 and IEC-60068-2-2
—
Storage
Temperature
–40ºC to +71ºC
Meets IEC-60068-2-1 and IEC-60068-2-2
—
Operating Relative
Humidity
10% to 90%, noncondensing
Meets IEC-60068-2-56
—
Storage Relative
Humidity
5% to 95%, noncondensing
Meets IEC-60068-2-56
—
Operating Shock 30 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with
MIL-PRF-28800F
—
Storage Shock 50 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with
MIL-PRF-28800F
—
Operating Vibration 5 Hz to 500 Hz, 0.31 grms
Meets IEC-60068-2-64
—
Storage Vibration 5 Hz to 500 Hz, 2.46 grms
Meets IEC-60068-2-64
Testprofile exceeds requirements of MIL-PRF-28800F,
ClassB
—
Altitude 2,000 m maximum (at 25 °C ambient temperature) —
Pollution Degree 2 —