NI PXI/PCI-5124 Specifications 26 ni.com
NI PCI-5124Note To ensure that the NI PCI-5124 cools effectively, make sure that the chassis in which
it is used has active cooling that provides at least some airflow across the PCI card cage.
To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty.
Refer to the Maintain Forced Air Cooling Note to Users included in the NI PCI-5124 kit
for important cooling information. The NI PCI-5124 is intended for indoor use only.
Specification Value Comments
Operating
Temperature
0 ºC to +45 ºC
Meets IEC-60068-2-1 and IEC-60068-2-2
—
Storage
Temperature
–40ºC to +71ºC
Meets IEC-60068-2-1 and IEC-60068-2-2
—
Operating Relative
Humidity
10% to 90%, noncondensing
Meets IEC-60068-2-56
—
Storage Relative
Humidity
5% to 95%, noncondensing
Meets IEC-60068-2-56
—
Storage Shock 50 g, half-sine, 11 ms pulse
Meets IEC-60068-2-27
Test profile developed in accordance with
MIL-PRF-28800F
—
Storage Vibration 5 Hz to 500 Hz, 2.46 grms
Meets IEC-60068-2-64
Testprofile exceeds requirements of MIL-PRF-28800F,
ClassB
—
Altitude 2,000 m maximum (at 25 °C ambient temperature) —
Pollution Degree 2 —