Figures 1 and 2 show the SCC-AO10 connected to a nonreferenced load and to a load with a high common-mode voltage.

E/M Series

DAQ Device SCC-AO10

3

 

AO (X )

 

2

Load

 

AO GND

 

1

 

I

 

Figure 1. SCC-AO10 Connected to a Nonreferenced Load

E/M Series

DAQ Device SCC-AO10

3

 

AO (X )

 

2

Load

 

AO GND

 

1

 

I

 

Figure 2. SCC-AO10 Connected to a Load with a High Common-Mode Voltage

Connecting Two SCC-AO10 Modules in Series

You can increase the maximum total output voltage to ±20 V by connecting two SCC-AO10 modules in series. Plug the two modules into sockets J17 and J18 of the SC-2345/2350, and wire the modules to a load as shown in Figure 3.