Maximum Working Voltage
(Signal +
Maximum working voltage refers to the signal voltage plus the CMV.
42.4 Vpeak or 60 VDC, | |
| Measurement Category I |
Caution This device is rated for Measurement Category I and is intended to carry signal
voltages no greater than 42.4 Vpeak or 60 VDC. Do not use this device for connection to signals or for measurements within Categories II, III, or IV.
Isolation Voltage
| |
Continuous ...................................... | 60 VDC, |
| Measurement Category I |
Withstand ........................................ | 2300 Vrms verified by a 5 s |
| dielectric withstand type test |
Environmental
Operating temperature............................ | 0 to 50 °C |
Storage temperature ............................... | |
Humidity ................................................ | 10 to 90% RH, noncondensing |
Maximum altitude .................................. | 2,000 m |
Pollution Degree (indoor use only)........ | 2 |
Stability
Recommended | 10 min (30 min for connected |
| E/M Series DAQ device) |
Output offset drift................................... | 300 μV/°C |
Gain temperature coefficient.................. | 300 ppm/°C |
© National Instruments Corporation | 9 |