Active mode of

Correction

 

measurement

Condition

parameter

(Std)

 

 

 

 

 

 

S, Sd, d

Distance

Install unit at 1 meter1

from flat target2

C, E

Azimuth and

Put unit in a level plane

and aim it to the North

 

elevation

direction3

 

 

Install unit at 1 meter1

dCE

Distance,

from flat target2

azimuth and

controlling its level

 

elevation

position and direction to

 

 

the North

1 – Accuracy of correction will depend on accuracy of installation against target. If distance is 0.9 meters, then unit will have 0.1 m offset during all future measurements.

2 – Target should not have high-reflective surface (white color is not considered high reflective in this case.

3 – Unit shall be well levelled with deviation from 0 not more than ±0.75 degrees. Horizontal direction shall be to Magnetic North (not True North). If devices detects large installation errors (unit is not well levelled or not aimed to the North), then LCD will show result “444” and “99”, which means that correction was not performed due to incorrect setup.

17

Page 21
Image 21
Newcon Optik LRB 6000CI operation manual Std Sd, d, Dce