OK International DX-250 specifications 5CONNECTING THE UNIT

Models: DX-250

1 15
Download 15 pages 51.01 Kb
Page 3
Image 3
5CONNECTING THE UNIT

5CONNECTING THE UNIT

CAUTION: A 5-micron filter (TSD800-6), not included, must be installed with the unit to ensure proper air filtration.

3

1

6.3Cycle Counter

The cycle counter records the numbers of dispense cycle being activated. Up to 65,535 cycles can be recorded. This number is shown at the lower right hand corner of the LCD. Press the Counter Reset button (3) to clear the counter.

7 TROUBLESHOOTING

2

4

Figure 2.0

Items

Description

1

To Air Source

2

Air Filter (Not Included)

3

Power Adapter

4

Foot Switch

6SETUP INSTRUCTIONS Refer to Figure 1.0

6.1Manual Dispense Cycle Setting:

1.Turn on the unit by pressing the Power button (1)

2.Press the Mode button (2) until “PURGE” appears on the Display

3.Turn up the air pressure by rotating the Air pressure regulator knob (6) until the desired pressure is indicated on the Air pressure gauge (5).

4.After filling the barrel or using pre-packaged Adhesives, attach syringe to receiver head assembly. Make certain that the syringe locks into place.

5.Connect the plug end of the receiver head assembly to one of the air dispense outlet (7)

6.If vacuum “suck back” is needed, rotate the Vacuum regulator (8) counter clockwise until the desired vacuum pressure is obtained

7.Press and hold the Footswitch to activate dispensed cycle. (The Manual Mode is now activated)

6.2Automatic Dispense Cycle Setting

1.Push the Mode button (2) to select “TIMED” mode

2.Rotate the Time control knob (9) to set desired dispense time

3.Repeat step 3 to 6 in Section 6.1

4.Depress Foot switch to activate timed dispensed cycle

PROBLEM

LCD does not light

System will not actuate

System will not pressurize

System will not pull vacuum

Inconsistent dispensing

POSSIBLE CAUSE

CORRECTION

No power input

Check power cord

 

connections

 

Turn on power

Foot switch not

Check foot switch

plugged in or

connection

improperly plugged

 

in

 

Defective foot switch

Foot switch needs to be

Broken wire or loose

repaired or replaced

Unplug power cord and

connection inside

disconnect air supply.

unit

Remove cover and check

 

for broken wires or loose

Defective solenoid

connections

Replace solenoid

Defective PC board

Replace PC board

 

 

Insufficient air

Increase air supply

pressure

pressure

Air hoses not

Check connection

plugged in

 

Regulator defective

Replace regulator

Vacuum setting is

Increase vacuum setting

too low

Replace solenoid valve

Defective solenoid

Replace venture

Defective vacuum

 

venturi

Replace needle valve

Defective vacuum

 

needle valve

 

Air bubbles in

Reduce vacuum setting

Adhesives

 

Dispense time is too

Increase dispensing tim

low

 

Dispense needle

Replace needle

started to clog

 

5

6

Page 3
Image 3
OK International DX-250 specifications 5CONNECTING THE UNIT