RS232 / RS422 WIRING
RS 232
COMPUTER HOOKUP:
RS 232: When connecting the unit to a computer with RS 232 communication, only three connections are needed. These connections are: Receive data, Transmit data and Ground. The connections should be made as follows:
RST | A | B | C | D | E |
ENTER LOCK | PRE A PRE B | VIEW | PRGM |
RECEIVE PIN (3)
TRANSMIT PIN (2)
GROUND PIN (5)
DP | COMPUTER |
Transmit data (pin 2) | Receive data |
Receive data (pin 3) | Transmit data |
Ground (pin 5) | Ground |
| RS 422 |
| RECEIVE A (+) PIN (3) |
| RECEIVE B |
RST A B C D E | TRANSMIT A (+) PIN (2) |
ENTER LOCK PRE A PRE B VIEW PRGM |
|
| TRANSMIT B |
| GROUND PIN (5) |
RS 422: When connecting the unit to a computer with RS 422,
five connections are needed. These connections are: Re- ceive data A (+), Receive data B
9 | 8 |
| 7 | 6 |
5 | 4 | 3 | 2 | 1 |
DP
COMPUTER
Receive data A(+) Receive data
Ground
RS 232 | RS 422 | ||
1. | INITIALIZE | 1. | INITIALIZE |
2. TRANSMIT | 2. TRANSMIT A (+) | ||
3. RECEIVE | 3. RECEIVE A (+) | ||
4. | N/C | 4. | N/C |
5. GROUND | 5. GROUND | ||
6. STROBE | 6. STROBE | ||
7. | N/C | 7. | TRANSMIT B |
8. | N/C | 8. | RECEIVE B |
9. | N/C | 9. | N/C |
PRINTER HOOKUP:
When connecting the unit to a printer, you must first program the desired baud rate, parity and strobe list with a computer. After the unit is programmed it can be connected to the printer. Connect the transmit line(s) of the unit to the receive line(s) of the printer and be sure that both devices have common grounds. When the strobe line is triggered the unit will transmit the selected strobe list which you had previously pro- grammed.
RS 232
STROBE PIN (6)
+12V PIN (7) (BOTTOM BOARD)
RST A B C D E | TRANSMIT PIN (2) |
ENTER LOCK PRE A PRE B VIEW PRGM |
|
GROUND PIN (5)
| RS 422 |
| STROBE PIN (6) |
| +12V PIN (7) |
| (BOTTOM BOARD) |
RST A B C D E | TRANSMIT A (+) PIN (2) |
ENTER LOCK PRE A PRE B VIEW PRGM |
|
| TRANSMIT B |
| GROUND PIN (5) |
19