FTB600B Series

5 Installation

Section 5 - Installing the FTB600B Series in the fluid line.

A. Make sure the fluid is compatible with PVDF (polyvinylidene fluoride) and meets viscosity, pressure and temperature parameters of the FTB600B Series. The fluid must also meet filtration requirements as listed in 3D.

B. Install the FTB600B series in the fluid line with the arrow pointing in the direction of the flow.

C. While installing the FTB600B Series in the fluid line be careful not to over-torque the end caps (on hose-barbed flowmeters) or other fittings on the flowmeter. Due to the relatively soft composition of PVDF the body or threads can be permanently distorted.

D. Attach wires to the readout display with the display and power off. Not only will this help to avoid a potential shock hazard, but it can also help prevent an error in hooking the flow meter to an incorrect 115 Vac supply.

E. Connect digital display to power supply and enter scaling factors for both the rate and total. Follow the manufacturers instructions for programming the digital display.

F. The FTB600B Series is now ready for use.

G. See Table 5-1 for straight pipe recommendations.