Omega FTB600B C. Recommended Viscosity, Range 1-5 cSt w/o correction, D. Filter Recommendations

Models: FTB600B

1 26
Download 26 pages 33.93 Kb
Page 9
Image 9
C. Recommended Viscosity

FTB600B Series

2

Operating Parameters

 

 

 

C. Recommended Viscosity

Range 1-5 cSt (w/o correction)

The effects of changing viscosity on the FTB600B Series are the same as any other turbine flow meter. It is important to remember that a turbine meter is a viscosity dependent device, where as the viscosity increases the linearity of the flow meter will decrease. (Water like viscosities are ideally suited for use with the FTB600B Series) The FTB600B Series is factory calibrated with water.

Correction procedure for higher viscosity

For viscosities greater than 5 cSt consult the factory. The FTB600B Series can be used on fluids greater than 5 cSt. however, the K-factor (linearity) will change. This requires a recalibration of the FTB600B Series at the known viscosity to determine the new KFactor.

D. Filter Recommendations

Meter

Micron

Mesh

FTB601B

35

400

FTB602B

50

300

FTB603B

100

80

FTB604B

100

80

FTB605B

100

80

FTB606B

100

80

5

Page 9
Image 9
Omega FTB600B C. Recommended Viscosity, Range 1-5 cSt w/o correction, Correction procedure for higher viscosity, Meter