HUMIDITY CALIBRATION KIT

HUMIDITY CALIBRATION KIT

The OMEGA® RH30-CAL Humidity Calibration Kit is needed to perform the humidity calibration of the RH31 Indicator and RH31-1/RH31-2 Probe assembly.

The Calibration Kit consists of two salt chambers held in a styrofoam base. One chamber contains a saturated salt solution of Lithium Chloride (to simulate 11% relative humidity). The other chamber contains a saturated salt solution of Sodium Chloride (to simulate 75% RH).

The RH31-3 cannot be field calibrated. No adapters are available.

INSTRUCTIONS TO DISASSEMBLE THE HUMIDITY PROBES

The potentiometer (pots) necessary for the humidity calibration procedures are located in the RH31-1/RH31-2 probes.

Each probe is slightly different, thus disassembly will vary. Perform the follow- ing procedures while referring to the corresponding steps pictured in the Figure for your particular probe type.

RH31 HUMIDITY/TEMPERATURE ADJUSTMENT

1.Please have RH30 Kit which OMEGA Engineering provide (30% RH).

2.Potential meter for adjustment is located about middle of handle portion of probe. (Only 1 potential meter)

3.As on Fig. 3 and 4, remove label covering potential meter.

Remove

LabelVR

Fig. 3. RH31-1 Probe

Remove

LabelVR

Fig. 4. RH31-2 Probe

4.Enter probe to calibration kit.

5.Turn potential meter to meet calibration kit reading. Once display is same or close to calibration kit, adjustment is done.

6.No calibration for temperature.