HUMIDITY CALIBRATION KIT
The OMEGA®
The Calibration Kit consists of two salt chambers held in a styrofoam base. One chamber contains a saturated salt solution of Lithium Chloride (to simulate 11% relative humidity). The other chamber contains a saturated salt solution of Sodium Chloride (to simulate 75% RH).
The
INSTRUCTIONS TO DISASSEMBLE THE HUMIDITY PROBES
The potentiometer (pots) necessary for the humidity calibration procedures are located in the
Each probe is slightly different, thus disassembly will vary. Perform the follow- ing procedures while referring to the corresponding steps pictured in the Figure for your particular probe type.
RH31 HUMIDITY/TEMPERATURE ADJUSTMENT
1.Please have RH30 Kit which OMEGA Engineering provide (30% RH).
2.Potential meter for adjustment is located about middle of handle portion of probe. (Only 1 potential meter)
3.As on Fig. 3 and 4, remove label covering potential meter.
Remove
LabelVR
Fig. 3. RH31-1 Probe
Remove
LabelVR
Fig. 4. RH31-2 Probe
4.Enter probe to calibration kit.
5.Turn potential meter to meet calibration kit reading. Once display is same or close to calibration kit, adjustment is done.
6.No calibration for temperature.