If the [E-##] or [Err] error messages are displayed, the unit will not function and must be returned to the factory for repair. (See section on Warranty for return instruction).

The following charts summarize the fault condition, and the associated indication.

High Target Temperature

Condition

 

Error

 

Measurement

Remedy

 

 

Code

 

Fault

 

 

 

 

 

 

 

 

 

 

 

 

 

Model OS951

 

[HI]

 

Target >550°F

 

Use higher

 

 

 

 

 

(287 °C)

 

OS950

 

 

 

 

 

 

 

 

Model OS953

 

[HI]

 

Target >1000

 

 

Use higher

 

 

 

 

 

°F (538 °C)

 

OS950

 

 

 

 

 

 

 

 

Model OS955

 

[HI]

 

 

Target >1207

 

Use higher

 

 

 

 

 

°F (653 °C)

OS950

 

 

 

 

 

 

 

Model OS956

 

[HI]

 

Target >1600

Contact

 

 

 

 

 

°F (871 °C)

Omega

 

 

 

 

 

 

 

 

 

 

Low Target Temperature

 

 

 

 

 

Condition

 

Error

 

Measurement

Remedy

 

 

Code

 

Fault

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model OS951

 

[LO]

 

Target < -50 °F

 

 

Contact

 

 

 

 

 

(-45 °C)

 

 

Omega

 

 

 

 

 

 

 

 

 

 

 

 

Model OS953

 

[LO]

 

Target <0 °F (-

 

Use lower

 

 

 

 

 

18 °C)

 

OS950

 

 

 

 

 

 

 

 

 

 

 

Model OS955

 

[LO]

 

 

Target <186 °F

 

 

Use lower

 

 

 

 

 

 

(86 °C)

 

 

OS950

 

 

 

 

 

 

 

 

 

 

 

 

Model OS956

 

 

[LO]

 

 

 

 

 

Use OS951 or

 

 

 

 

Target <186 °F

 

 

 

 

 

 

 

(86 °C)

OS953

 

 

 

 

 

 

 

 

 

 

 

 

Other Errors

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Condition

 

Error

 

Measurement

Remedy

 

 

 

Code

 

Fault

 

 

 

 

 

 

 

 

 

 

 

 

 

 

High Ambient

 

[HI A]

 

Ambient > 122

 

 

Move to

Temperature

 

 

 

 

°F (50 °C)

 

 

alternate

 

 

 

 

 

 

 

 

 

location

 

 

 

 

 

 

 

 

 

 

Low Ambient

 

 

 

 

Ambient < 32

 

 

 

 

 

 

[LO A]

 

 

 

Move to

Temperature

 

 

 

 

°F (0 °C)

 

alternate

 

 

 

 

 

 

 

 

location

 

 

 

 

 

 

 

 

 

Low Battery

 

[N/A]

 

 

Battery power

 

 

Replace

 

 

 

 

 

 

low

 

 

battery

 

 

 

 

 

 

 

 

 

 

 

Dead Battery

 

 

 

 

 

Dead battery

 

 

 

 

 

 

[

----]

 

 

 

 

 

Replace

 

 

 

 

 

 

 

 

 

battery

 

 

 

 

 

 

 

 

 

Processing

 

[E-##]

 

 

Internal fault

 

 

 

Repeat

Error

 

or [Err}

 

 

 

 

 

measurement,

 

 

 

 

 

 

 

if

error

 

 

 

 

 

 

 

 

 

remains,

 

 

 

 

 

 

 

 

 

contact

 

 

 

 

 

 

 

 

 

Omega

 

 

 

 

 

 

 

 

 

 

 

 

8

Page 13
Image 13
Omega OS950 manual High Target Temperature