If the
The following charts summarize the fault condition, and the associated indication.
High Target Temperature
Condition |
| Error |
| Measurement | Remedy | ||||||
|
| Code |
| Fault |
|
|
|
|
| ||
|
|
|
|
|
|
|
| ||||
Model OS951 |
| [HI] |
| Target >550°F |
| Use higher | |||||
|
|
|
|
| (287 °C) |
| OS950 | ||||
|
|
|
|
|
|
|
| ||||
Model OS953 |
| [HI] |
| Target >1000 |
|
| Use higher | ||||
|
|
|
|
| °F (538 °C) |
| OS950 | ||||
|
|
|
|
|
|
|
| ||||
Model OS955 |
| [HI] |
|
| Target >1207 |
| Use higher | ||||
|
|
|
|
| °F (653 °C) | OS950 | |||||
|
|
|
|
|
|
| |||||
Model OS956 |
| [HI] |
| Target >1600 | Contact | ||||||
|
|
|
|
| °F (871 °C) | Omega | |||||
|
|
|
|
|
|
|
|
|
| ||
Low Target Temperature |
|
|
|
|
| ||||||
Condition |
| Error |
| Measurement | Remedy | ||||||
|
| Code |
| Fault |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
Model OS951 |
| [LO] |
| Target < |
|
| Contact | ||||
|
|
|
|
|
|
| Omega | ||||
|
|
|
|
|
|
|
|
|
|
|
|
Model OS953 |
| [LO] |
| Target <0 °F (- |
| Use lower | |||||
|
|
|
|
| 18 °C) |
| OS950 | ||||
|
|
|
|
|
|
|
|
|
|
| |
Model OS955 |
| [LO] |
|
| Target <186 °F |
|
| Use lower | |||
|
|
|
|
|
| (86 °C) |
|
| OS950 | ||
|
|
|
|
|
|
|
|
|
|
|
|
Model OS956 |
|
| [LO] |
|
|
|
|
| Use OS951 or | ||
|
|
|
| Target <186 °F |
|
| |||||
|
|
|
|
| (86 °C) | OS953 | |||||
|
|
|
|
|
|
|
|
|
|
|
|
Other Errors |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
Condition |
| Error |
| Measurement | Remedy |
| |||||
|
| Code |
| Fault |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| |||
High Ambient |
| [HI A] |
| Ambient > 122 |
|
| Move to | ||||
Temperature |
|
|
|
| °F (50 °C) |
|
| alternate | |||
|
|
|
|
|
|
|
|
| location | ||
|
|
|
|
|
|
|
|
|
| ||
Low Ambient |
|
|
|
| Ambient < 32 |
|
|
|
|
| |
| [LO A] |
|
|
| Move to | ||||||
Temperature |
|
|
|
| °F (0 °C) |
| alternate | ||||
|
|
|
|
|
|
|
| location | |||
|
|
|
|
|
|
|
|
| |||
Low Battery |
| [N/A] |
|
| Battery power |
|
| Replace | |||
|
|
|
|
|
| low |
|
| battery | ||
|
|
|
|
|
|
|
|
|
|
| |
Dead Battery |
|
|
|
|
| Dead battery |
|
|
|
|
|
| [ |
|
|
|
|
| Replace | ||||
|
|
|
|
|
|
|
|
| battery | ||
|
|
|
|
|
|
|
|
| |||
Processing |
|
|
| Internal fault |
|
|
| Repeat | |||
Error |
| or [Err} |
|
|
|
|
| measurement, | |||
|
|
|
|
|
|
| if | error | |||
|
|
|
|
|
|
|
|
| remains, | ||
|
|
|
|
|
|
|
|
| contact | ||
|
|
|
|
|
|
|
|
| Omega | ||
|
|
|
|
|
|
|
|
|
|
|
|
8