Omega Speaker Systems HH303 I. Introduction, II. Specifications, Numerical Display, Resolution

Models: HH303

1 14
Download 14 pages 55.88 Kb
Page 4
Image 4
I. Introduction:

Type K J Thermometer

I. Introduction:

This instrument is a digital thermometer for use with any K-type and J-type thermocouple as temperature sensor.

Temperature indication follows National Bureau of Standards and IEC584 temperature/voltage table for K-type and J-type thermocouples.

II.Specifications:

Numerical Display:

4 digital liquid crystal display

Measurement Range:

 

 

 

 

K-Type:

-200°C ~ 1370°C

-328°F ~ 2498°F

 

J-Type:

-200°C ~ 760°C

-328°F ~ 1400°F

 

Resolution:

 

 

 

 

 

K-Type:

-200°C~ 800°C

0.1°C;

800°C ~1370°C

1°C

J-Type:

-328°F~ 1000°F 0.1°F;

else

1°F

 

-200°C~ 600°C

0.1°C;

600°C ~760°C

1°C

 

-328°F~ 1000°F 0.1°F; else

1°F

 

Maximum Voltage at Thermocouple Input:

 

 

60V DC, or 24Vrms AC

Environmental:

οOperating Temperature and Humidity:

0°C ~50°C (32°F ~ 122°F) ; 0 ~ 80% RH

οStorage Temperature and Humidity:

-10°C to 60°C (14°F ~ 140°F); 0 ~ 80% RH

οAltitude up to 2000 meters.

Accuracy: at ( 23 ± 5°C )

 

Range

Accuracy

K-Type -200°C ~ 1370°C

±(0.1% reading + 0.7°C)

J-Type -2000°C ~ 760°C

±(0.1% reading + 0.7°C)

K-Type -328°F ~ 2498°F

±(0.1% reading + 1.4°F)

J-Type -328°F ~ 1400°F

±(0.1% reading + 1.4°F)

For T1-T2 Measurement, the accuracy is ±( 0.2% T1-T2 reading + 1.7°C) or ±( 0.2% T1-T2 reading + 3.4°F)

￿

Page 4
Image 4
Omega Speaker Systems HH303 manual I. Introduction, II. Specifications, Numerical Display, Measurement Range, Resolution