11.- ADJUSTMENT AND CALIBRATION PROCEDURE

11.1.- RTD MODELS

A decade box or RTD simulator is required.

1.- Check that the instrument is not powered.

2.- Connect the RTD simulator or decade box to the signal input connector (See paragraph 8.3).

3.- Remove the rear cover taking off all the screws to obtain access to the control board.

(See page 7 paragraph 7).

4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to start the adjust.

5.- Adjust the RTD simulator to generate the low signal indicated in TABLE-A.

6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1 located on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.

7.- Adjust the RTD simulator to generate the high signal, see TABLE-B.

8.- Display must shows the DISPLAY value indicated in TABLE-B, if not turn the trimmer PT2 located on the Signal Conditioner (See paragraph 10.1) until display shows the DISPLAY value.

9.- Repeat steps 5 to 8 until the two values are correct.

10.- Check that the linearity remains correct, testing some other points of the range.

11.2.- THERMOCOUPLE MODELS

A Thermocouple simulator with Cold Junction compensation is required.

1.- Check that the instrument is not powered.

2.- Connect the Thermocouple simulator to the signal input connector (See paragraph 8.4).

3.- Remove the rear cover taking off all the screws to obtain access to the control board. (See page 7 paragraph 7).

4.- Power up the instrument with the appropriate power supply and wait until 3 minutes before to

start the adjust.

5.- Adjust the Thermocouple simulator to generate the low signal indicated in TABLE-A.

6.- Display must shows the DISPLAY value indicated in TABLE-A, if not turn the trimmer PT1located on the Signal Conditioner (See paragraph 10.2) until display shows te correct value.

7.- Adjust the Thermocouple simulator to generate the high signal indicated in TABLE-B.

8.- Display must shows the DISPLAY value, if not turn the trimmer PT2 located on the Signal Conditioner (See paragraph 10.2) until obtain the DISPLAY value.

9.- Repeat steps 5 to 8 until the low and the high signals are correct.

10.- Check that the linearity remains correct, testing some other points of the range.

TABLE - A

MODEL

LOW SIGNAL

DISPLAY

 

 

 

 

 

 

M1

0°C

000.0

M2

0°F

0000

M3

0°F

000.0

J1

0°C

0000

J2

32°F

0032

K1

0°C

0000

K2

32°F

0032

T1

0°C

0000

T2

0°F

0000

E1

0°C

0000

E2

32°F

0032

R1

300°C

0300

R2

32°F

0032

S1

200°C

0200

S2

32°F

0032

L1

0°C

0000

 

 

 

TABLE - B

MODEL

HIGH SIGNAL

DISPLAY

 

 

 

 

 

 

M1

600°C

600.0

M2

500°F

0500

M3

200°F

200.0

J1

600°C

0600

J2

1000°F

1000

K1

1200°C

1200

K2

2000°F

2000

T1

400°C

0400

T2

400°F

0400

E1

600°C

0600

E2

1200°F

1200

R1

1700°C

1700

R2

3100°F

3100

S1

1700°C

1700

S2

3100°F

3100

L1

600°C

0600

 

 

 

Page : 10