4. Error Messages and Troubleshooting
Display | Cause | Solution |
|
|
|
| Air plug is disconnected. | Insert the air plug securely. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Arm cuff is applied too loosely. | Apply the arm cuff tighter. |
| Refer to section 3.1. | |
|
| |
|
|
|
| Air is leaking from the arm cuff. | Replace the arm cuff with a new one. |
| Refer to section 5.3. | |
|
| |
|
|
|
|
| Repeat measurement. Remain still and do not |
|
| talk during measurement. |
| Movement during measurement and the arm | Refer to section 3.3. |
|
| |
| cuff has not been inflated sufficiently. | If “E2” appears repeatedly, inflate the arm cuff |
|
| manually until it is 30 to 40 mmHg above your |
|
| previous measurement result. |
|
| Refer to section 3.3. |
|
|
|
| The arm cuff was inflated above 299 mmHg | Do not inflate the arm cuff above 299 mmHg. |
| when inflating the arm cuff manually. | Refer to section 3.3. |
|
|
|
|
| Repeat measurement. Remain still and do not |
| Movement during measurement. | talk during measurement. |
|
| Refer to section 3.3. |
|
|
|
|
| Remove any clothing interfering with the arm |
| Clothing is interfering with the arm cuff. | cuff. |
|
| Refer to section 3.1. |
| Device error. | Contact your OMRON retail outlet or |
| distributor. | |
|
| |
|
|
|