Troubleshooting

If a problem occurs while you are using the unit, please check the following information before calling.

The nebulization rate is extremely low, or the unit does not nebulize at all.

Where to Inspect

How to correct

 

 

Replace the batteries with new ones.

Is the Battery Low Indicator turned on?

(Refer to Page 10)

Clean the Mesh Cap by disinfecting to remove

any stains.

(Refer to Page 18)

• If the trouble is not corrected by the above

Is the Mesh dirty?

mentioned procedure, replace the Mesh cap with a new one.

(Refer to Page 19)

The Power Indicator does not turn on and the unit does not nebulize.

Troubleshooting Troubleshooting

Where to Inspect

How to correct

 

 

 

Are the polarities of batteries or rechargeable

Insert batteries or rechargeable batteries in the

correct alignment.

 

batteries placed in the wrong direction?

 

 

(Refer to Page 10)

 

 

 

 

Are the batteries worn?

Replace the batteries with new ones.

 

(Refer to Page 10)

 

 

 

 

Is the capacity of the rechargeable batteries

Recharge the batteries with the commercially

available recharger or replace them

(NiMH) low?

with new ones.

 

 

 

 

 

 

Is the AC Adapter securely plugged into

 

 

the electric outlet or the connection port

Insert it correctly.

(Refer to Page 11)

of Main Unit?

 

 

 

 

 

Is the Medication Bottle installed to the Main

Install it correctly.

(Refer to Page 13)

Unit correctly?

 

 

 

 

 