ERROR INDICATORS AND TROUBLESHOOTING TIPS

ERROR INDICATORS

SYMBOL

 

CAUSE

CORRECTION

 

 

 

 

 

 

 

Air plug disconnected.

Insert the plug securely. Refer to “Applying the

 

 

Arm Cuff”.

 

 

 

 

 

 

 

 

 

Arm cuff not applied correctly.

Apply the arm cuff correctly. Refer to “Applying

 

the Arm Cuff”.

 

 

 

 

 

 

 

 

 

Air is leaking from the arm cuff.

Replace the cuff with the new one.

 

 

 

 

 

 

 

 

Repeat measurement. Remain still and do not talk

 

 

 

during measurement. Refer to “Taking a

 

Cuff over inflated.

Measurement”.

 

 

If “E2” appears repeatedly, inflate the cuff manually

 

 

 

 

 

 

until it is 30 to 40 mmHg above your previous

 

 

 

measurement result. Refer to “Taking a Measurement”.

 

 

 

 

 

 

The arm cuff was inflated above 299

Do not inflate the cuff above 299 mmHg. Refer to

 

mmHg when inflating the cuff manually.

“Taking a Measurement”.

 

 

 

 

 

 

 

 

Repeat measurement. Remain still and do not talk

 

Movement during measurement.

during measurement. Refer to “Taking a

 

 

 

Measurement”.

 

Clothing is interfering with the

Remove any clothing interfering with the arm cuff.

 

arm cuff.

Refer to “Applying the Arm Cuff”.

Device error.

Contact Customer Service.

Cuff is not applied correctly.

Refer to “Applying the Arm Cuff”.

Batteries are worn

Replace the four batteries. Refer to “Battery Installation”.

39

Page 39
Image 39
Omron Healthcare BP760 instruction manual Error Indicators and Troubleshooting Tips