MEMORY FEATURE INSTRUCTIONS

1. How to store the measured blood pressure values...

The measured and displayed blood pressure and pulse rate readings are automatically stored in the memory.

The memory can store the readings of up to fourteen measurements.

The memory stores the most recent 14 readings. When more than 14 readings have been taken the earliest readings are erased and replaced by the most recent readings. The memory is not erased when the batteries are replaced.

2. How to recall previously measured readings from memory...

While the () symbol is displayed, you can recall and display readings from the memory starting with the previous measured readings.

When you push the MEMORY Button “M”, if there is any measured reading stored in the memory, M1 indicating the latest memory, the systolic and the diastolic blood pressures, and the pulse rate are displayed alternatively.

In order to recall the readings measured previously, push the MEMORY Button “M” again. The values will be displayed as M2, M3,... etc.

If there is no measured reading in the memory, a blank screen with dashed lines will be displayed, when the MEMORY Button “M” is pushed.

3. How to erase the memory...

All the measured readings stored in the memory can be erased simultaneously by operating the buttons as follows:

While the () symbol is displayed, push the MEMORY “M” and the START Buttons simultaneously. Hold both buttons down for at least two seconds, a blank screen with dashed lines will be displayed and the measured readings stored in the memory will be erased.

TROUBLESHOOTING

Error Indicator

 

Cause

Correction

 

 

 

 

 

 

 

 

 

 

 

Cuff under-inflated

Retake measurement by setting

 

 

 

 

 

 

cuff inflation level 40 mmHg

 

 

 

 

 

 

higher than previous level, by

 

 

 

 

 

 

keep pressing the START

 

 

 

 

 

 

BUTTON until the cuff

 

 

 

 

 

 

pressure reaches the

 

 

 

 

 

 

necessary level.

 

 

 

 

 

 

 

 

 

 

 

Movement during

Be sure not to move and talk

 

 

 

 

measurement.

during the measurement.

 

 

 

 

 

 

 

 

 

 

 

No pulse or blood pressure

Wrap the cuff snugly around

 

 

 

 

your wrist, carefully read and

 

 

 

 

reading due to very weak

 

 

 

 

repeat steps listed under “How

 

 

 

 

pulse.

 

 

 

 

To Apply The Wrist Cuff”

 

 

 

 

 

 

 

 

 

 

 

 

(page 6).

 

 

 

 

 

 

 

 

 

 

 

Cuff over-inflated (more

Release the START BUTTON

 

 

 

 

than 300 mmHg).

before the pressure indication

 

 

 

 

 

 

reaches 280 mmHg or higher.

 

 

 

 

 

 

 

 

 

 

 

 

 

Replace both “AAA” batteries

 

 

 

 

Battery voltage is

 

 

 

 

excessively low

 

(See page 5).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTES:

-Wait 2-3 minutes before taking another measurement.

-If you continue to have difficulty obtaining a reading, call

OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350. Please tell the representative

you have an Omron HEM-629 Portable Wrist Blood Pressure Monitor.