Omron Healthcare HEM-907XL manual Blood Pressure in Manual Mode

Models: HEM-907XL

1 17
Download 17 pages 14.75 Kb
Page 13
Image 13

HOW TO MEASURE

BLOOD PRESSURE (IN MANUAL MODE)

1. Push the ON/OFF (power) Button to turn on the power.

2. Set the MODE Selector to "MANU".

3.Set the P-SET (inflation level setting) Knob to "AUTO" or the target pressure value.

4. Measure the patient’s arm size and wrap appropriate cuff over the patient's arm. (Refer to pages 14 and 15.)

HOW TO MEASURE

BLOOD PRESSURE (IN MANUAL MODE)

5.Place the stethoscope on the patient's arm.

6. Push the START Button to start the measurement.

• Do not push the START Button without wrapping the cuff.

• Do not squeeze or press the cuff during the measurement.

• If you want to inflate again after the start of deflation, push the START Button.

• If you want to accelerate deflation after the start of deflation, push the DEFLATION (deflation control) / Measurement Results Display Switch Button. Each time the Button is pushed, cuff is deflated rapidly in increments of 5 to 10 mmHg.

7. Take the readings.

8. Push the STOP Button to remove air inside the cuff.

• The unit does not automatically deflate in the MANU Mode.

9. Push the ON/OFF (power) Button to turn off the power.

If the monitor determines that the pressure value is not correct, an error display appears (Er1 to 9). In this case, refer to Page 28 and start the measurement again.