FEATURES AND FUNCTIONS

Names of the Parts

FEATURES AND FUNCTIONS

Display

Main unit

1. Display (See Page 11 for detail)

SYS

XL

 

2. HIDE (non-display) Button

 

DIA

mmHg

 

6.

ON/OFF (power)

 

Button

7.

START Button

Diastolic blood pressure

Systolic blood pressure

Ready to Measure / Pulse synchronization

Pulse level /

Number of irregular pulses

DC8V

3.DC jack

4.P-SET (pressure setting) Knob

PULSE

 

P-

 

140

SET

100

180

 

 

 

 

 

220

AUTO

 

260

 

 

280

mmHg

 

BPM

AC

 

 

 

HIDE

 

MODE

VG. SINGLE

A

 

 

 

 

MANU.

 

 

CHECK

ON/OFF

 

 

 

START

A

D

E

 

 

 

 

 

 

 

 

 

V

 

F

L

A

 

 

G

 

 

 

 

TIO N

 

./

 

 

 

 

 

 

1

st

/2

nd/3rd

 

 

 

 

 

 

 

 

 

 

 

STOP

 

8.

DEFLATION

 

(deflation control)

 

/Measurement

 

Result Display

 

Switch Button

9. Air Connector

10. STOP Button

SYS

mmHg

DIA

mmHg

PULSE

Number of irregular pulse waves: The pulse wave having a difference of more than 25% from the average interval of pulse waves is determined irregular. Number of irregular pulse waves is displayed up to three times.

Example of display

 

 

No. of irregular pulses

 

1

2

3

5. MODE Selector

1.Display: Displays blood pressure and, pulse rate readings, and oscillation pulse level.

2.HIDE Button: Switches display and non-display of measured results.

3.DC jack: Connects the AC adapter.

4.P-SET (pressure setting) Knob: In the AUTO position, inflation level is automatically set. Otherwise, inflation level can optionally be set manually between 100 and 280 mmHg.

5.MODE Selector: Selects the operation mode.

One-time Measurement Mode (SINGLE Mode): Measurement with automatic inflation.

Average Mode (AVG Mode): Automatically measures two (or three) times consecutively.

Auscultation Mode (MANU Mode): Automatic inflation, automatic deflation, and pressure display for auscultation (does not measure blood pressure).

Check Mode (CHECK Mode): Checks the accuracy of pressure display. Displays only pressure.

6.ON/OFF (power) Button: Turns on or off the unit.

7.START Button: Starts the measurement.

8.DEFLATION (deflation control) /Measurement Result Display Switch Button:

In the MANU Mode, deflates the cuff by approximately 5 to 10 mmHg with each push during deflation.

In the AVG Mode, switches the display of average values and the measurement results with each push.

9.Air Connector: Connects the air tube.

10.STOP Button: Stops the measurement and deflates air rapidly.

Contents when the AVG Mode is selected

AVG. : Mean value

1st : First reading

2nd : Second reading

3rd : Third reading

BPM AC

HIDE

Pulse rate

Battery level

Displays for the operable level.

Blinks

for a low charged

 

level

Displays for the inoperable level.

Charging

Displays when the battery pack is being charged.

External power source

Displays AC when the unit is connected to the external power source via the AC adapter.