Omron Healthcare NE-U17 Parts that can be disinfected by boiling, Fill the cooling water

Models: NE-U17

1 24
Download 24 pages 25.71 Kb
Page 10
Image 10

How to Prepare the Unit before Inhalation

– Clean the parts before starting the inhalation.

[Parts that can be disinfected by boiling]

(For the disinfecting method, refer to Page 17.)

Submerge these parts in a container filled with sufficient water and boil for 10 minutes.

Medication cup cover

Washer

Inhalation hose (M)

with a cuff

How to remove the washer

Place the medication cup cover on its side and insert a small item (such as a paper clip) into the hole of the cover.

Pinch and remove the washer gently.

How to attach the washer

Hold the medication cup cover and place the washer securely with your fingertips as illustrated in the figures on the left.

Make sure the washer is placed flat and not upside down.

1 Filling the cooling water.

1)Remove the nebulisation section by rotating the medication cup cover-fixing lever clockwise.

1

2

 

 

water

 

level

2) Fill the cooling water.

Fill the cooling water up to the “water level” mark that specifies the water level in the water tank. The proper amount is approximately 375 ml.

water level

[Parts that cannot be disinfected by boiling]

(For the care and disinfection, refer to Page 17.)

 

Air duct cover

Mouthpiece

Air filter case

set

Air filter

 

 

Fan cover

 

Medication cup holder

Medication cup

Fan

Warning

Be sure to immediately dry the cleaned and disinfected parts, then store them in a dry place in order to prevent them from getting contaminated in the future.

water level

If the water level in the water tank is lower than the “water level” mark, a “lower water level in water tank” error will be displayed.

If the water temperature in the water tank is low, the nebulisation rate may drop. (Recommended temperature is approximately 26°C (78.8 °F).

Replace water in the water tank each time the unit is used.

When the unit is used for a long time continuously, replace water at least every twenty-four hours.

(For continuous use or long periods, refer to Page 13.)

9

Page 10
Image 10