(4)Condenser lens astigmatism correction
-Press STIG to activate stigmator control.
-Press COND soft key so that it is highlighted (the default is OBJ).
-Condense the beam.
-Use MULTIFUNCTION knobs to correct the astigmatism. The beam should be round when changing the setting of the INTENSITY knob, or the filament image will be sharpest if filament is
-To see the sharp image of filament, three conditions,
-Press STIG to
(4)Pay attention to filament image while slowly increase filament current to saturate (prefer slightly
(5)Check aperture center by spreading the beam, make sure that beam spreads without sweep.
(C)Alignment of Imaging System:
(i)Voltage Center Alignment
To align the imaging system, the AC component is superimposed on the accelerating voltage in order to move the image radially. The center of the radial movement (called “voltage center”) is then matched with the center of the fluorescent screen. The main purpose of this step is to center the beam tilt so there won’t be a tendency to have a
(1)
-Rough focus the sample at 10,000 to 20,000X.
-Find an identifiable object and center it on the screen
-Tilt the sample. This may cause the center of the field of view to move towards the edge of the projection screen if the specimen is not at eucentric height position.
-Set the specimen height at the point where there is no movement of the specimen upon tilting by adjusting the z control (screw on the goniometer just below the sample holder).
-Bring the tilt mechanism to 0o.
(Hint: if autofocus is way off,
(2)Pivot point X and Y
-Move to center of sample (i.e., any hole or away from the sample).
-Press PIVOT PT X soft key in ALGN.
-Condense beam.
-Use MULTIFUNCTION knobs to bring two circles together.
-Press PIVOT PT Y soft key in ALGN.
-Condense beam.
-Use MULTIFUNCTION knobs to bring two circles together.
7