15.
6If “Er.2” is shown on the display,the operating range
is too low.
Solution: Area/Room is too cold for thermometer to
take a reading.Move to another area within the 61ºF to
104ºF (16.0ºC – 40.0ºC) range and retake temperature.
Allow 30 minutes for unit to adjust to the room
temperature.
7If “Er.3” is shown on the display,there is a system error.
Solution: Remove the battery and replace it in order
to reset.
ASSISTANCE
For assistance or the Authorized Norelco Service
Location nearest you,call toll free:
1-877-SEE TEMP(733-8367)
or visit our website:www.sensortouch.com
Authorized Norelco Service Location information is
available 24 hours a day,7 days a week.
14.
TROUBLESHOOTING
WARNING MODES
1Low Battery Warning
When the display flashes the temperature reading and
a blinking battery symbol,the 9V battery is low and
should be replaced.See Battery Installation section for
instructions on installing/replacing the battery.
2If“– –” and a blinking batter y symbol appear on the display,
there is not enough energy in the battery to measure a
temperature.Replace the battery. See Battery Installation
section for instructions.
3If a flashing “107.6ºF” is shown on the display,the target
temperature is too high for thermometer to capture
a reading.
Solution: This would indicate a person’s temperature
is above 107.6ºF (42.0ºC),and a physician should be
contacted immediately.
4If “- -” is shown on the display,the target temperature
is too low for thermometer to capture a reading.
Solution: This would indicate a person’s temperature
is below 96ºF (35.5ºC),and a physician should be
contacted immediately.
5If “Er.1” is shown on the display,the operating range
is too high.
Solution: Area/Room is too warm for thermometer
to take a temperature.Move to another area within
the 61ºF to 104ºF (16.0ºC – 40.0ºC) range and retake
temperature.Allow 30 minutes for unit to adjust to
the room temperature.
+-
+-
.1