Philips MML1801 user manual ABC Selection, Transfer Function, Advanced, Linear, Monitor TG3

Models: MML1801

1 28
Download 28 pages 45.25 Kb
Page 19
Image 19

6.3.5ABC Selection

1) Press Enter button when the ABC Selection is highlighted on Image Enhancement The ABC Selection window appears.

ABC Selection

ABC Off

ABC Reduced

ABC Full

2)Press the Adjust + or Adjust - button until the desired item is highlighted.

- ABC off

Manual back light control regulated by panel button from 250 cd/m² to 500 cd/m²

- ABC full

Automatic control of back light, the back light is function of ambient light and follow a semilogaritmic courve from 200 cd/m² (at 1 lux) to 500 cd/m² (at 1000 lux).

-ABC reduced

The same as ABC full but with full scale of 350 cd/m² (at 1000 lux).

3)Press the Enter button to confirm your selection and return to Image Enhancement window

6.3.6Transfer Function

1) Press Enter button when the Transfer Functiopn is highlighted on Image Enhancement The Transfer Function window appears.

Transfer Function

TG3 - PMS

Linear

DICOM

Monitor TG3

2)Press the Adjust + or Adjust - button until the desired item is highlighted. Indicate the output curve relative to the input signal:

- TG3 - PMS

Fixed output curve PMS like.

- Linear

The Tranfer Function is transparent

- DICOM

The output curve follows the DICOM curve

- Monitor TG3

The output curve is the same as PMS standard monitor.

3)Press the Enter button to confirm your selection and return to Image Enhancement window

6.3.7Advanced

1)Press Enter button when the Adavanced is highlighted on Main Menu The Advanced window appears.

19

MML1801

Page 19
Image 19
Philips MML1801 user manual ABC Selection, Transfer Function, Advanced, Linear, Monitor TG3