Component Description
ITEM | DESCRIPTION |
FAir Inlet Fitting
A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose.
GAlarm
An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply.
HManifold Outlet (Optional) Manifold with 3 primary outlets.
I | Rear Slide Mount with dove tail. |
Installation
WARNING
•Read this User Manual before installing or operating the
•Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor.
1.Secure the
2.It is recommended to install a condensation trap in the air supply line.
3.Connect the air and oxygen supply lines to the appropriate inlet fittings on the bottom of the Blender.
4.Attach a flowmeter, or other metering device to one of the outlet ports.
Primary Outlets Flow capacity:
•High Flow Blender (PM 5200 Model) 15 lpm to 120 lpm
•Low Flow Blender (PM 5300 Model) 3 lpm to 30 lpm
Auxiliary Outlet:
This will bleed off some of the air/oxygen mixture to maintain concentration accuracy at the Low Flow setting.
•High Flow Blender (PM 5200 Model) 15 lpm or less
•Low Flow Blender (PM 5300 Model) 3 lpm or less
5.Attach a supply line to the outlet port of the flowmeter.