Raynger 3i Series Operator’s Manual 2-21

7. Set the TAM value by pressing the andbuttons.

8. Press the ACTIVATE button to turn on the ambient temperature compensation

function. The Tamb icon will be displayed indicating that it is active. (Pressing

ACTIVATE again turns off the function.)

9. Pull trigger and reset the emissivity to the proper value for the target. You can

now take the target’s temperature using normal procedures. The current tempera-

ture and all computed values, as well as the analog and digital outputs, will now

be based on the compensated measurement.

Note: TAM only affects readings in the RUN loop; LOG loop readings are not affected.

2.4.7.2 SETUP Values for LOG

Figure 2-11 illustrates SETUP for the LOG loop.

To setup values for the LOG modes and functions, do the following:

1. Release or unlock the trigger, if necessary.

2. Press the SET button. The SET icon will be activated.

3. Press the RUN/LOG button, if necessary, so that the LOG icon is activated.

4. Press the or buttons to select the LOG location number (LOC mode).

MODE
RUN
LOG
SET
MODE
RECALL
ACTIVATE
LOCK
LASER
LOG RCL SET °C °F
HI ( ) LO
( ) Tamb
Press the and
buttons to change the
Log Location (LOC)
number.
You can select and
log data in any location
number from 1 to 100.
Press the Set button
again, if the display
goes out, to continue
the SETUP functions
RUN
LOG
SET
MODE
RECALL
ACTIVATE
LOCK
LASER
LOG RCL SET °C °F
HI ( ) LO
( ) Tamb
Turn High Alarm
on or off with the
Activate button Change the High Alarm
temperature value with
the and buttons

LOC Mode HAL Mode

TRIGGER

Released & Unlocked

BUTTON

SET & LOG

LOOP ICON

LOG SET
RUNLOGRECALL
MODEACTIVATE
LOCK
LASER
RUNLOGRECALL
MODEACTIVATE
LOCK
LASER
Figure 2-11: SETUP Loop–LOG Values