GB

Disposal

The Papillon nasal mask, AeroFix II headgear, and AeroFix plus tubing clip contain no toxic substances. Dispose of the nasal mask in accordance with applicable national laws and regulations. Except where stipulated otherwise, before being taken to a recycling center the nasal mask should be cleaned in accordance with the procedures described in this Instruction Manual.

Warranty

Atwo-year warranty applies for manufacturing defects. The warranty does not cover normal wear and tear. For further details, see the MAP Medizin-Technologie GmbH General Terms and Conditions of Business.

MAP Medizin-Technologie GmbH shall not be liable for any impairments to the safety, reliability, or performance of the product caused by:

use other than that specified in the Instruction Manual

use in conjunction with other products not designed for this purpose

any technical alterations made to the product.

The mask must always be cleaned and disinfected before being returned for servicing. On completion of the above procedure, dry the mask and pack it in a securely fastened transport bag.

Symbols

See Instruction Manual

Warning! Failure to follow this instruction could pose a health risk.

REF Part number

Lot number

Date of manufacture (month, year)

Washing temperature 30 °C

Do not use chlorine bleach

Do not iron

Do not spin-dry

Technical Data

Standards

CE labeling in accordance with EC

directive 93/42/EEC, class IIa Quality Management: development,

manufacture, sale, and service certified in accordance with ISO 13485:2003 quality management principles.

EN ISO 17510-2

(Sleep apnea respiratory therapy, Part 2: masks and application accessories)

Validated Cleaning and Disinfection

The general suitability of the Papillon® nasal mask for effective cleaning and disinfection by the procedures specified in this Instruction Manual has been established by an independent accredited test laboratory.

Flow Resistance

Pressure drop measured with mask cushion size 2 at 50 l/min = max. 0.1 hPa

at 100 l/min = max. 0.4 hPa

Acoustic Pressure

30.8dB(A), measured with artificial head at 10 hPa according to EN ISO 17510-2

Tubing Connection

22 mm for elastomer tubing

Weight

Approx. 92 g

Therapeutic Pressure

4 to 20 hPa

Materials Used

 

GmbH

Mask (latex-free):

silicone, PC

 

Headgear (neoprene-free): nylon, lycra, spandex,

Technologie-

 

polyurethane, polyester

 

 

Cleaning of Headgear

 

 

Pressure/Flow Curve

 

Medizin

 

 

Mean flow [l/min]

 

 

(6) Copyright 2005 by MAP

 

P [hPa]

11.2005

Pressure Units

 

1 hPa = 1 mbar = approx. 0.98 cm H2O

 

 

 

 

 