GB
Disposal
The Papillon nasal mask, AeroFix II headgear, and AeroFix plus tubing clip contain no toxic substances. Dispose of the nasal mask in accordance with applicable national laws and regulations. Except where stipulated otherwise, before being taken to a recycling center the nasal mask should be cleaned in accordance with the procedures described in this Instruction Manual.
Warranty
A
MAP
•use other than that specified in the Instruction Manual
•use in conjunction with other products not designed for this purpose
•any technical alterations made to the product.
The mask must always be cleaned and disinfected before being returned for servicing. On completion of the above procedure, dry the mask and pack it in a securely fastened transport bag.
Symbols
See Instruction Manual
Warning! Failure to follow this instruction could pose a health risk.
REF Part number
Lot number
Date of manufacture (month, year)
Washing temperature 30 °C
Do not use chlorine bleach
Do not iron
Do not
Technical Data
Standards
CE labeling in accordance with EC
directive 93/42/EEC, class IIa Quality Management: development,
manufacture, sale, and service certified in accordance with ISO 13485:2003 quality management principles.
EN ISO
(Sleep apnea respiratory therapy, Part 2: masks and application accessories)
Validated Cleaning and Disinfection
The general suitability of the Papillon® nasal mask for effective cleaning and disinfection by the procedures specified in this Instruction Manual has been established by an independent accredited test laboratory.
Flow Resistance
Pressure drop measured with mask cushion size 2 at 50 l/min = max. 0.1 hPa
at 100 l/min = max. 0.4 hPa
Acoustic Pressure
30.8dB(A), measured with artificial head at 10 hPa according to EN ISO
Tubing Connection
∅22 mm for elastomer tubing
Weight
Approx. 92 g
Therapeutic Pressure
4 to 20 hPa
Materials Used |
| GmbH | |
Mask | silicone, PC | ||
| |||
Headgear | Technologie- | ||
| polyurethane, polyester | ||
|
| ||
Cleaning of Headgear |
|
| |
Pressure/Flow Curve |
| Medizin | |
|
|
Mean flow [l/min]
|
| (6) Copyright 2005 by MAP |
| P [hPa] | 11.2005 |
Pressure Units |
| |
1 hPa = 1 mbar = approx. 0.98 cm H2O |
| |
|
| |
|
| REF 560 185 |