8Fully unscrew and remove the dial then pull the forehead support out of the mask frame (Fig. C-7).

9Remove the forehead support pads from the forehead support (Fig. C-8).

10 Squeeze and push the upper side tabs on the cushion clip out of the mask frame. Pivot away from the mask, and gently pull out (Fig. C-9).

11 Separate the cushion from the cushion clip (Fig. C-10).

12 Remove the ports cap from the mask frame (Fig. C-11).

Cleaning the Mask in the Home

Notes:

The mask and headgear can only be cleaned by handwashing.

The headgear may be washed without removing the headgear clips.

WARNING

Do not use aromatic-based solutions or scented oils (eg, eucalyptus or essential oils), bleach, alcohol or products that smell strongly (eg, citrus) to clean any of the mask components. Residual vapours from these solutions can be inhaled if not rinsed thoroughly. They may also damage the mask, causing cracks.

Do not iron the headgear as the material is heat sensitive and will be damaged.

Daily/After Each Use

To optimize the mask seal, facial oils should be removed from the cushion before use.

Handwash the mask by gently rubbing in warm (30°C) water using mild soap or diluted dishwashing detergent.

Do not twist the valve during cleaning as this may damage or tear the valve membrane or hinge. Take care when handling the valve.

If the vent requires cleaning use a soft bristle brush.

Rinse all components well with drinking quality water and allow them to air dry out of direct sunlight. Do not soak the mask.

Weekly

Handwash the headgear in warm (30°C) water using mild soap or diluted dishwashing detergent. Rinse well and allow it to air dry out of direct sunlight.

4

Page 8
Image 8
ResMed Mirage Quattro manual Cleaning the Mask in the Home, Daily/After Each Use, Weekly