Using humidifiers
When a device integrated with the HumidAire 2i™ or H4i™ humidifier is not connected to the mains power supply, the humidifier will only operate in passive unheated mode despite the
S9™ and H5i™ devices operate normally in active heated mode when connected to the battery or to the mains power supply.
Cleaning and maintenance
WARNING
Do not immerse the RPS II in water, and do not use liquids to clean any part of the product.
1.Disconnect the battery from the power supply unit and the device. Remove all cables.
2.Wipe the exterior of the battery with a clean cloth.
Storing
CAUTION
Batteries must be charged to 100% indicated by four green LEDs and switched off before storing.You must recharge the batteries to 100% again after three months of storage. When not in use, all
Store the RPS II in a cool, dry place.
Note: A fully charged battery when left switched on,
Servicing
The RPS II is intended to provide safe and reliable operation when used and maintained in accordance with the instructions provided by ResMed. There is no servicing necessary during the lifetime of the battery.
The service life of the RPS II exceeds 500 recharge cycles. After this cycle, the battery will last about 60% of the time from its original condition. A full charge in an older battery will not last as long as in a new battery. ResMed recommends that the RPS II be tested for battery duration regularly. As with all electrical equipment, if any irregularity becomes apparent, you should exercise caution and contact an authorised ResMed service representative.
Traveling
Consult with your carrier if you intend to take the battery on board with your device.
4