Problem/Possible Cause | Solution |
Displays error message: Exxxx Call Service (where xxxx defines an error) | |
Component failure. | Record error number and contact your ResMed service centre. |
Displays error message: HIGH LEAK!!! Adjust Mask
You have experienced excessively high leak levels for more than 20 seconds.
Check that your air tubing is connected properly. Adjust headgear.
The following message is displayed on the LCD after you try to update settings or copy data to the Data Card: Card Error Remove Card
Data Card is not inserted correctly.
You may have removed the Data Card before settings were copied to the VPAP device.
Ensure that the Data Card is inserted with the arrow facing up as far as it can go.
Reinsert the Data Card and wait for the Settings Success Remove Card or Copy Complete Remove Card message to appear on the LCD.
The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Invalid Remove Card
The identification details on the | Contact your clinician/service provider immediately. |
Data Card do not match the
details on your device.
The following message is displayed on the LCD after you try to update the settings using the Data Card: Settings Error Remove Card
There is a data error on the Data | Contact your clinician/service provider immediately. |
Card. |
|
The following message is NOT displayed on the LCD after you try to update the settings using the Data Card: Settings Success Remove Card
The settings were not updated. | Contact your clinician/service provider immediately. |
14