OPERATING INSTRUCTIONS
Alarm, safety function, and self diagnostic function
This unit has the alarm and safety functions shown in Table 2, and also self diagnostic function.
Table 2 Alarm and safety functions |
| ||
Kind of alarm | Situation | Indication | |
or safety | |||
|
| ||
High | If the upper or lower |
| |
temperature | temp. sensor senses a | Alarm lamp flashes. | |
alarm | temperature ≧6oC. | ||
Low | If the upper or lower |
| |
temperature | temp. sensor senses a |
| |
alarm | temperature ≦2oC. |
| |
| If the | Unit should already be | |
thermostat sensor | |||
in alarm. No further | |||
(high) alarm | senses a temperature | ||
| approximately>10oC. | indication required. | |
Power failure | Interruption of power to | Alarm lamp flashes. | |
unit or power switch is | |||
alarm | Temp. indication off. | ||
turned off. | |||
|
| ||
Door alarm | Outer door is ajar or | Door check lamp is lit. | |
open. | |||
|
| ||
|
|
| |
| No key operation on | Monitor bottle temp. is | |
Auto return | control panel for ≧90 | ||
displayed. | |||
| seconds. | ||
|
| ||
| Upper temperature | Alarm lamp flashes. | |
| Monitor bottle temp. | ||
| sensor goes open or | ||
| E1 are alternately | ||
| short circuit. | ||
| displayed. | ||
|
| ||
| Lower temperature | Alarm lamp flashes. | |
| Monitor bottle temp. | ||
| sensor goes open or | ||
| E2 are alternately | ||
Thermal | short circuit. | ||
displayed. | |||
sensor |
| ||
| Alarm lamp flashes. | ||
abnormality |
| ||
Defrost sensor goes | Monitor bottle temp. | ||
| |||
| open or short circuit. | E3 are alternately | |
|
| displayed. | |
| Alarm lamp flashes. | ||
| Monitor bottle temp. | ||
| open or short circuit. | E4 are alternately | |
|
| displayed. |
Buzzer
Intermittent tone.
Unit should already be in a high temp. alarm.
Intermittent tone.
Intermittent tone after a delay of 2 minutes.
Intermittent tone
Safety operation
Remote alarm is active.
Remote alarm is active. Compressor ON. Heater OFF.
Remote alarm is active.
Not affected.
Calibration mode is canceled.
Remote alarm is active.
Note: The remote alarm contacts will activate in conjunction with the audible alarm. The remote alarm contacts can be deactivated while pressing the BUZZER key when the unit is in alarm, except in a power failure alarm where the contacts will remain in an alarm condition until the power failure is corrected. All alarm functions will
Alarm set procedure
Press the ALARM TEST key. The audible and visual alarm indicators will activate along with the remote alarm contacts. The alarm condition will remain active for 90 seconds and then terminate automatically. Otherwise the alarm test can be terminated manually at any time during the rest by pressing the ALARM TEST key a second time.
18